Oxygen is always the first thing on and the last thing off.
Using an Oxygen Tank — Startup
1. Open the oxygen tank. On an industrial or medical tank, turn the top valve counterclockwise to open. On
an O2Ready prefilled tank, turn the small knob on the side of the regulator to open.
2. Set your desired flow rate on the regulator.
3. Power on the ozone generator, set your desired concentration, and begin therapy.
Shutdown
1. Power off the ozone generator.
2. Let oxygen continue flowing for about 30 seconds to purge ozone from the system and tubing.
3. If using a bubbler, disconnect the tubing from the ozone output port to prevent water or oil from
backflowing.
4. Close the oxygen tank by turning the valve clockwise.
5. Turn the flow rate to 1 LPM and watch the needle drop into the red or refill zone. This relieves trapped
pressure and purges remaining ozone from the system and tubing.6. Return the flow rate to 0 (off). If using a prefilled tank, remove the regulator from the tank.
Using an Oxygen Concentrator — Startup
1. On the Promolife low flow regulator's Oxygen In port, make sure the OFF portion of the 3 way valve is
pointed toward the regulator.
2. Turn the concentrator on and set it to 5 LPM. Let it run for 5 to 10 minutes so oxygen purity can rise, then
set it to 2 LPM.
3. Turn the 3 way valve so the OFF portion points to the open, perpendicular position (the side with nothing
attached).
4. Adjust the low flow regulator to your desired flow rate.
5. Power on the ozone generator, select your desired concentration, and begin therapy.
Shutdown
1. Power off the ozone generator.
2. Let oxygen continue flowing for about 30 seconds to purge ozone from the system and tubing.
3. If using a bubbler, disconnect the tubing from the ozone output port to prevent water or oil from
backflowing.
4. Turn off the oxygen concentrator.
5. Reset the 3 way valve so the OFF portion points toward the regulator, ready for next use.